Evaluation of ultrasound parameters used as a diagnostic method for stress urinary incontinence

PURPOSE: to verify the sensitivity, specificity and accuracy of some ultrasonographic measurements in the diagnosis of stress urinary incontinence (SUI). METHODS: transvaginal ultrasound measurements of the bladder neck descent, urethral mobility and urethrovesical funneling caused by urination effort were performed in 40 women with SUI and in 40 women from a control group. Age, parity and the number of pregnancies were different in both groups. Several cut points were performed to evaluate the sensitivity, specificity and accuracy of these measurements as a diagnostic tool for SUI. The urodynamic results were used as gold standard. Univariated analysis was done using Yates chi2 Test and Pearson chi2 Test. RESULTS: in the best cut point for bladder neck descent measurements, sensitivity was 40%, specificity was 72% and accuracy was 57%; in the best cut point for urethral mobility measurements, sensitivity was 40%, specificity was 70% and accuracy was 55%; in the best cut point for urethrovesical funneling measurements, sensitivity was 58%, specificity was 48% and accuracy was 52%; in the best cut point for the addition of the differences of these three measurements, sensitivity was 32%, specificity was 62% and accuracy was 48%. CONCLUSION: vaginal ultrasonography was not a valid diagnostic method for stress urinary incontinence in the present study.